A gravity-fed, multi-filter system left no visible thrombus in the filtrate, and filtered porcine and human blood showed no sign of red blood cell damage on blood smears.
Visible in the final filtrate after clotted bovine blood and saline were passed through the filters.
Seen on blood smears or filtered porcine and human blood. No red cell fragments (schistocytes) were found.
Unchanged after filtering a human blood sample (PT 10.5, INR 0.93 before and after).
Retriever Medical is developing a disposable, gravity-fed filtration system intended to filter a person's own (autologous) whole blood before it is returned to them. Three chambers in a row each hold a removable filter: 2mm, 250 microns and 40 microns.
This early proof-of-concept study asked three questions: does the filter remove clot, does it damage red blood cells, and does it change how the blood clots?


The final filtrate had no visible thrombus or particles. In the pig blood, potassium, calcium and hematocrit did not differ between the unfiltered and filtered samples, and smears showed no change in red cell shape. In the human blood, smears showed no red cell or platelet injury, and most markers stayed the same.
Two values differed. LDH was higher in the filtered human sample (323 vs 171 U/L), and PTT was higher in the unfiltered samples (52 vs 31). The authors note more blood samples need to be analyzed, and that further studies are needed on the safety of reinfusing blood filtered this way.

Preview the first page below, or download the PDF (6 pages, about 0.8 MB).
In live porcine, the ClotHound spheres were expanded 20-40% beyond vessel diameter and moved back and forth without significant damage to the vessel lining.
The 24F aspiration/guide catheter (7.1 mm ID) tracked through the serial 90-180 degrees turns, with no lumen collapse at a 6 mm radius of curvature.
The expandable wire sphere pushed on the vessel wall with at most 22 N of force, far below the force reported to rupture the saphenous vein and pulmonary artery.
A gravity-fed multi-filter system removed 100% of test particles from a blood analogue, at more than 500 cc per minute.